| Bone and Anatomical Region |
Femur, tibia, humerus, or other long-bone applications |
Matches the implant geometry to the natural shape, canal anatomy, and loading pattern of the target bone. |
The nail should be selected specifically for the involved bone and its intended entry point. |
Nails are not universally interchangeable between bones. Entry-point anatomy, canal curvature, and nearby soft tissues must be assessed. |
| Fracture Location |
Proximal, shaft, distal, or segmental fracture |
Determines how much nail support and how many locking points are needed around the fracture. |
Longer nails and increased fixation options may be useful when the fracture extends toward an end of the bone or includes multiple fragments. |
Very proximal or distal fractures may require a nail with specially positioned locking holes and angularly stable fixation options. |
| Fracture Pattern |
Transverse, short oblique, long oblique, spiral, comminuted, or segmental |
Helps determine the level of mechanical stability required from the nail and locking screws. |
Interlocking nails are commonly used for many diaphyseal and selected metaphyseal long-bone fractures, especially when control of length and rotation is important. |
Highly unstable, comminuted, or segmental patterns may require careful locking configuration, reduction control, and sometimes supplementary techniques. |
| Nail Length |
Short, standard, or long nail, selected by measurement |
Provides adequate fracture-span coverage and distributes load along the medullary canal. |
A nail should generally extend across the fracture and terminate in a safe, well-supported region without creating joint conflict. |
Length is evaluated using imaging, intraoperative measurement, and the position of the entry point. Excessive length or prominence can cause complications. |
| Nail Diameter |
Diameter chosen according to the medullary canal and implant system |
Balances structural strength, canal fit, and preservation of bone blood supply. |
A larger diameter may improve bending stiffness when safely accommodated; a smaller diameter may be needed in a narrow or irregular canal. |
The selected diameter must account for canal measurements, reaming strategy, fracture stability, and the risk of cortical damage. |
| Reamed or Unreamed Technique |
Reamed nail or unreamed nail |
Creates an appropriate canal environment and allows selection of a suitable implant diameter. |
Reamed fixation is frequently used in many adult long-bone shaft fractures; unreamed techniques may be considered in selected situations. |
Choice depends on fracture characteristics, soft-tissue condition, bone quality, associated injuries, and surgeon judgment. Reaming can affect intramedullary pressure and must be performed carefully. |
| Locking Screw Configuration |
Static locking, dynamic locking, or combined locking |
Controls rotation and axial movement while maintaining alignment and length. |
Static locking is commonly considered for unstable, comminuted, or length-unstable fractures. Dynamic options may permit controlled axial compression in selected patterns. |
Locking mode should match the fracture's stability. Incorrect configuration may lead to shortening, malrotation, delayed union, or excessive implant stress. |
| Number of Locking Points |
One or more screws on each side of the fracture |
Improves control of rotation, translation, and axial loading. |
More fixation points may be useful in short distal or proximal segments, poor bone quality, or unstable fracture patterns. |
The number and position of screws should preserve available bone, avoid joint penetration, and provide an appropriate balance between stability and construct flexibility. |
| Locking Hole Design |
Static round holes, dynamic slots, multiplanar holes, or angle-stable holes |
Provides different options for controlling movement and directing screw placement. |
Multiplanar or angle-stable options can be helpful when the available bone segment is short or when additional control of a metaphyseal fragment is needed. |
The hole pattern must be compatible with the intended screws and targeting instruments. Screw trajectory should be confirmed radiographically. |
| Entry Point and Nail Curvature |
Bone-specific entry point with straight, bowed, or anatomically contoured nail |
Facilitates insertion while reducing malalignment and cortical stress. |
The nail curvature should correspond to the patient's anatomy and the selected entry approach. |
An incorrect entry point or mismatch between nail curvature and anatomy can contribute to malalignment, cortical perforation, or difficulty with insertion. |
| Material and Surface |
Implant-grade stainless steel or titanium alloy; polished or treated surface |
Provides biocompatibility, fatigue resistance, and suitable mechanical behavior. |
Material selection is generally based on the implant system, required mechanical properties, imaging considerations, and patient-specific factors. |
Material stiffness, elastic modulus, corrosion resistance, and compatibility with other implants should be considered. Do not mix components unless compatibility is confirmed. |
| Bone Quality |
Normal bone, osteopenic bone, or osteoporotic bone |
Influences screw purchase, fixation strength, and the risk of loss of alignment. |
Patients with reduced bone density may benefit from a construct with suitable screw distribution and enhanced fixation options. |
Bone quality should be assessed with imaging and clinical history. Implant selection alone cannot compensate for inadequate reduction or poor fixation technique. |
| Open or Closed Fracture |
Closed fracture or open fracture with varying soft-tissue injury |
Places the implant decision within the broader context of contamination, soft-tissue coverage, and infection risk. |
Intramedullary fixation may be used in selected open fractures after appropriate irrigation, debridement, antibiotics, and soft-tissue management. |
Open fractures require urgent multidisciplinary assessment. The fixation plan must be coordinated with infection prevention and soft-tissue treatment. |
| End-Cap and Nail Prominence |
Flush, recessed, or slightly prominent end-cap options |
Secures the nail, helps manage nail position, and may reduce soft-tissue irritation. |
An end-cap may be selected to adjust nail depth or facilitate later removal when clinically indicated. |
Prominence near a joint or tendon can cause pain and irritation. Final nail depth should be checked clinically and with imaging. |
| Targeting and Instrumentation |
Standard targeting jig, radiographic freehand locking, or computer-assisted guidance |
Supports accurate placement of interlocking screws and reduces operative uncertainty. |
The method depends on nail design, fracture location, available equipment, and surgical expertise. |
Instrument compatibility, screw length measurement, and confirmation of screw position are essential. All components should be checked before use. |
| Postoperative Objective |
Early mobilization, protected weight bearing, or staged loading |
Aligns the mechanical construct with the rehabilitation plan and biological healing response. |
A load-sharing nail may support earlier functional rehabilitation in appropriately stabilized fractures. |
Weight-bearing status must be individualized according to fracture stability, reduction quality, bone condition, healing progress, and the treating surgeon's protocol. |